Before you get started … • 51
English
4.5 Connecting the power supply
The head slide is supplied with power from batteries.
The batter holder supplied already contains 4 Mignon
batteries, type AA, 1.5 volt. To connect the power
supply for the head slide, proceed as follows:
1. Remove the cover of the battery compartment.
2. Remove the connection cable from the battery
compartment.
3. Connect the battery holder to the connection cable.
4. Insert the battery holder into the battery
compartment.
ATTENTION!
Damage to device and malfunction due to
clamped or pinched cable
Lay the battery cable before closing the
battery compartment so that it cannot get
clamped between the cover and the battery
compartment.
5. Slide the cover onto the battery compartment until
it audibly latches into position.
4.6 Calibrate head slide
Before you can measure height using the device for the
first time, you have to calibrate it. You do so by
“teaching” the device a reference measurement.
Automated calibration You can perform automated calibration using the
calibration rod supplied.
1. Press the Start key of the head slide.
“----” appears in the display.